Radiation doses during computed tomography scan imaging in Hassan II Hospital, Agadir, Morocco
Slimane Semghouli1, Bouchra Amaoui2, Soukaina Wakrim2
1Higher Institute of Nursing Professions and Health Techniques, Agadir, Morocco.
Introduction:
the aim of this study was to establish local Diagnostic Reference Levels (DRLs) for four adult Computed Tomography (CT) examinations in the radiology department, Hassan II hospital of Agadir.
Methods:
during this survey, we have examined the data of 200 patients at an average of 50 per localization. A General Electric 16 CT with automatic exposure control system was used to perform all CT examinations. Scanner acquisition parameters, number of series, contrast medium use, rotation time plus slice thickness, the displayed Computed Tomography Dose Index (CTDIvol), and the Dose Length Product (DLP) were among the data collected for each diagnostic exam chosen. To evaluate the DRL and effective dose, the conversion factor and formalism of the International Commission on Radiological Protection (ICRP) were utilized.
Results:
the average effective dose (Eeff), the displayed CT dose index (CTDIvol), dose length product (DLP) were (7.28±2.35) mSv, (10.80±3.80) mGy and (428.35±138.26) mGy.cm respectively at chest CT. For abdomen-pelvis CT scan, there were (12.48±5.58) mSv, (9.30±2.99) mGy and (805.43±359.98) mGy. Those at chest abdomen-pelvic CT scan were (11.72±3.98) mSv, (10.82±2.53) mGy and (755.97±251.52) mGy.cm respectively. For lumbar CT, there were (12.12±2.32) mSv, (26.46±5.24) mGy and (787.00±149.37) mGy.cm respectively.
Conclusion:
the findings of this study shows that our values are slightly higher than those of developed countries. This first-ever CT practice evaluation at Agadir's Hassan II Hospital reinforced the need for further radiology training for computed tomography practitioners on parameters influencing image quality, dose, and protocols improvement.
More Related Videos
10:33Expedited Radiation Biodosimetry by Automated Dicentric Chromosome Identification ADCI and Dose Estimation
Published on: September 4, 2017
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation I: X-ray and CT
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...