The impact of Morse taper implant design on microleakage at implant-healing abutment interface
Soyeon Kim1, Joo Won Lee1, Jae-Heon Kim1
1Department of Oral Anatomy and Dental Research Institute, School of Dentistry, Seoul National University.
Abstract:
The microbial leakage at the implant-abutment interface (IAI) is one cause of peri-implant infection that puts long-term implant stability at risk. The present study compared two types of Morse taper implants in terms of sealing performance at the implanthealing abutment interface. Three implant systems, one of which exhibited a partial face-contact design (TSO) and the other two a line-contact design (TSM and BLT), were analyzed in vitro using two quantitative implant leakage measuring techniques: air-injection pressure measurement test and microbial examination. An in-house-developed device was used to determine implant leakage by recording the initial drop in pressure while injecting air through the implant. The microbial examination measured the absorbances of culture mediums in which the inoculated implants were contained. Significant difference was found between the partial face- and line-contact groups (p<0.001). Both tests revealed that partial face-contact implants are more prone to leakage than line-contact implants.