Measurement of Heart Contractility in Isolated Adult Human Primary Cardiomyocytes
Published on: August 9, 2022
Crack Sensing of Cardiomyocyte Contractility with High Sensitivity and Stability
Li Wang1,2, Xingyuan Xu1,2, Jun Chen1,2
1School of Mechanical Engineering, Qilu University of Technology (Shandong Academy of Sciences), Jinan 250353, China.
Abstract:
Measuring myocardial contractility is of great value in exploring cardiac pathogenesis and quantifying drug efficacy. Among the biosensing platforms developed for detecting the weak contractility of a single layer of cardiomyocytes (CMs), thin brittle metal membrane sensors with microcracks are highly sensitive. However, their poor stability limits the application in long-term measurement. Here, we report a high stability crack sensor fabricated by deposition of a 105 nm thick Ag/Cr with microcracks onto a carbon nanotubes-polydimethylsiloxane (CNT-PDMS) layer. This brittle-tough bilayer crack sensor achieved high sensitivity (gauge factor: 108 241.7), a wide working range (0.01-44%), and high stability (stable period >2 000 000 cycles under the strain caused by a monolayer of CMs). During 14-day continuously monitoring CMs culturing and drug treatment testings, the device demonstrated high sensitivity and stability to record the dynamic change caused by contractility of the CMs.
More Related Videos
06:42Author Spotlight: Real-Time Measurements of Calcium and Contractility Parameters in Human Induced Pluripotent Stem Cell-Derived Cardiomyocytes
Published on: May 26, 2023
08:29Assessment of Myofilament Ca2+ Sensitivity Underlying Cardiac Excitation-contraction Coupling
Published on: August 1, 2016