Performance evaluation of presepsin using a Sysmex HISCL-5000 analyzer and determination of reference interval
Taewon Kang1,2, Jeaeun Yoo3, Hyunyu Choi3
1Departement of Laboratory Medicine, Seoul St. Mary's Hospital, College of Medicine, The Catholic University of Korea, Seoul, South Korea.
Background:
Analytical evaluation of newly developed presepsin by a Sysmex HISCL-5000 (Sysmex, Japan) automated immune analyzer was performed.
Methods:
For evaluation, sepsis patient samples were collected before treatment in an emergency department. Precision, linearity, limit of blank/limit of detection, method comparisons, and reference intervals were evaluated. Method comparisons were performed using a PATHFAST immune analyzer (LSI Medience Corporation, Japan).
Results:
Precision using a 20x2x2 protocol for low (306 pg/mL) and high (1031 pg/mL) levels resulted in within-laboratory standard deviation (95% confidence interval [CI]) and coefficient of variation (CV) %, which were as follows: 15.3 (13.1-18.7), 5.5% and 47.7, (40.5-58.1), 6.4%, respectively. Linearity using patient samples and calibrators were measured from 201 to 16,177 and 188 to 30,000 pg/mL, respectively. The regression equation was y = -23.2 + 1.008x (SE = 162.4) for low levels and y = 779.9 + 1.006x (SE = 668) for high levels. Method comparison by Passing-Bablock analysis was as follows: y = -209.77 + 1.047x (Syx = 335.3). The correlation coefficient (95% CI) was 0.869 (0.772-0.927) with statistical significance (p < 0.001). Reference intervals from 120 normal healthy subjects showed that 300 pg/mL was the cut off. Presepsin tended to show a higher value at higher ages and in males. Presepsin showed correlation with some parameters, and the correlation coefficient (p value) were as follows: hematocrit, 0.198 (0.03); eGFR (CKD-EPI), -0.240 (0.0129); MDRD-eGFR, -0.194 (0.048), respectively.
Conclusion:
Presepsin measurement by HISCL-5000 showed reliable performance. Further clinical studies are required for the diagnosis and prognosis of sepsis.