Scatter estimation for a digital radiographic system using convolution filtering.

L A Love, R A Kruger

    Medical Physics
    |March 1, 1987
    PubMed
    Summary
    This summary is machine-generated.

    Related Concept Videos

    You might also read

    Related Articles

    Articles linked to this work by shared authors, journal, and citation graph.

    Sort by
    Same author

    Pathologic changes in the cardiac interstitium of mice infected with encephalomyocarditis virus.

    Cardiovascular pathology : the official journal of the Society for Cardiovascular Pathology·2015
    Same author

    Time resolved imaging through a highly scattering medium.

    Applied optics·2010
    Same author

    New fourier transform all-reflection interferometer.

    Applied optics·2010
    Same author

    Denture cream: an unusual source of excess zinc, leading to hypocupremia and neurologic disease.

    Neurology·2008
    Same author

    Breast cancer in vivo: contrast enhancement with thermoacoustic CT at 434 MHz-feasibility study.

    Radiology·2000
    Same author

    Approaches for identifying and defining environmentally associated rheumatic disorders.

    Arthritis and rheumatism·2000
    Same journal

    Correction to "On the shape of the radiation survival curve in tumor spheroids: The role of oxygen heterogeneity".

    Medical physics·2026
    Same journal

    Multi-view constrained semi-supervised vertebra detection for 3D ultrasound spine volume.

    Medical physics·2026
    Same journal

    Accuracy of quantitative <sup>177</sup>Lu SPECT/CT imaging: A systematic review.

    Medical physics·2026
    Same journal

    Physics-constrained dual-domain network for CBCT reconstruction from orthogonal X-rays in gynecologic radiotherapy.

    Medical physics·2026
    Same journal

    Decomposition-based harmonization for quantitative PET imaging across scanners and radiotracers.

    Medical physics·2026
    Same journal

    Development and evaluation of an in vivo dose-based monitoring system for electron FLASH radiation therapy.

    Medical physics·2026
    See all related articles

    A new convolution-filtering method effectively estimates scatter in digital subtraction angiography (DSA) images. This technique uses specific 2D exponential kernels to reduce scatter, improving image quality with minimal error.

    Area of Science:

    • Medical Imaging
    • Image Processing
    • Radiology

    Background:

    • Digital subtraction angiography (DSA) is crucial for visualizing blood vessels.
    • Image scatter degrades the diagnostic quality of DSA images.
    • Accurate scatter estimation is essential for robust DSA image analysis.

    Purpose of the Study:

    • To investigate a convolution-filtering method for scatter estimation in DSA images.
    • To identify optimal convolution kernels for scatter reduction.
    • To develop a practical method for scatter correction in DSA.

    Main Methods:

    • Evaluated over 175 convolution kernels on phantom DSA images (head, chest, pelvis).
    • Utilized 2D exponential kernels with varying Full Width at Half Maximum (FWHM).

    Related Experiment Videos

  • Applied a novel method using lead beam stops and kernel convolution for scatter correction.
  • Main Results:

    • 2D exponential kernels with 50-150 pixels FWHM best reproduced scatter fields.
    • Achieved root-mean-square percentage errors between 4% and 8%.
    • A 75-pixel FWHM kernel yielded an average error of 6.6% across diverse cases.

    Conclusions:

    • The proposed convolution-filtering method accurately estimates and corrects for scatter in DSA.
    • The technique is implementable using lead beam stops and convolution kernels.
    • This approach enhances DSA image quality for improved clinical interpretation.