The potential 'blue light hazard' from LED headlamps
Richard B Price1, Daniel Labrie2, Braden Sullivan1
1Department of Dental Clinical Sciences, Division of Prosthodontics, Faculty of Dentistry, Dalhousie University, 5981 University Ave., P.O. BOX 15000, Halifax, Nova Scotia B3H 4R2, Canada.
Abstract:
Many dental personnel use light-emitting diode (LED) headlamps for hours every day. The potential retinal 'blue light hazard' from these white light headlamps is unknown.
Methods:
The spectral radiant powers received from direct and indirect viewing of an electronic tablet, an LED curing light, a halogen headlamp, and 6 brands of LED headlamps were measured using integrating spheres attached to fiberoptic spectroradiometers. The spectral radiant powers were measured both directly and indirectly at a 35 cm distance, and the maximum daily exposure times (tMAX) were calculated from the blue weighted irradiance values.
Results:
The headlamps emitted very different radiant powers, emission spectra, and color temperatures (K). The total powers emitted at zero distance ranged from 47 mW from the halogen headlamp to 378 mW from the most powerful LED headlamp. The color temperatures from the headlamps ranged from 3098 K to 7253 K. The tMAX exposure times in an 8 h day when the headlamps were viewed directly at a distance of 35 cm were: 810 s from the halogen headlamp, 53 to 220 s from the LED headlamps, and 62 s from the LED curing light. Light from the LED headlamps that was reflected back from a white reference tile 35 cm away did not exceed the maximum permissible exposure time for healthy adults. Using a blue dental dam increased the amount of reflected blue light, but tMAX was still greater than 24 h.
Conclusions:
White light LED headlamps emit very different spectra, and they all increase the retinal 'blue light hazard' compared to a halogen source. When the headlamps were viewed directly at a distance of 35 cm, the 'blue light hazard' from some headlamps was greater than from the LED curing light (tMAX = 62 s). Depending on the headlamp brand, tMAX could be reached after only 53s. The light from the LED headlamps that was reflected back from a white surface that was 35 cm away did not exceed the maximum permissible ocular exposure limits for healthy adults.
Clinical Relevance:
Reflected white light from dental headlamps does not pose a blue light hazard for healthy adults. Direct viewing may be hazardous, but the hazard can be prevented by using the appropriate blue-light blocking glasses.
More Related Videos
10:41Enhanced Electron Injection and Exciton Confinement for Pure Blue Quantum-Dot Light-Emitting Diodes by Introducing Partially Oxidized Aluminum Cathode
Published on: May 31, 2018
09:29In Vivo Investigation of Antimicrobial Blue Light Therapy for Multidrug-resistant Acinetobacter baumannii Burn Infections Using Bioluminescence Imaging
Published on: April 28, 2017
Related Concept Videos
Biasing of P-N Junction
In equilibrium, no external voltage is applied across the p-n junction. The depletion region is formed at the junction interface due to the diffusion of carriers, which leaves behind charged dopants, acceptors on the p-side, and donors on the n-side. These immobile charges create an electric field that prevents further diffusion of carriers. The related energy band...
Electrical Energy
Photoreceptors and Visual Pathways
Survey Safety