Related Experiment Video
Updated: Sep 3, 2025

Author Spotlight: Development of a Standardized Acupuncture Tool Inspired by Advanced Techniques for Improved Safety and Precision
Published on: January 10, 2025
[Design, manufacture and evaluation of a multi-parameter controllable automatic fire-acupuncture instrument]
Xiao-Qin Wang1, Wei Xu1, Xiao-Feng Han1
1Department of Dermatology, Chongqing First People's Hospital/Chongqing Hospital of Chinese Medicine, Chongqing 400011, China.
Abstract:
A multi-parameter controllable automatic fire-acupuncture instrument was developed by integrating traditional fire needling with modern medical device technology. A gun-like appearance was designed for easy hand-held operation, the electromagnetic induction was for heating needle body, a scale knob was for controlling the needle insertion depth, the combination of electromagnetic ejection and spring return was for the precise control of the needle retention time; and the changeable single ste-rile needle or multiple needles were adopted to meet individual demand, obtain high efficiency and prevent infection. All of these designs are associated with the overall process control system to ensure the exact controllability of needle body temperature, needling density, insertion depth and needle retention time. Besides, this device is advantageous at handy and aseptic operation with high efficiency, conformability and visualization. In this research, this instrument was tested in animals for the impacts of automatic fire needling on skin damage and fur growth. It is found that the accurate control of each parameter is of the significant advantage in the safety and effectiveness of treatment, which lays a solid foundation for the subsequent systematic review on safety and effectiveness.

