Photon-counting detector CT and energy-integrating detector CT for trabecular bone microstructure analysis of cubic
Benjamin Klintström1, Lilian Henriksson2,3, Rodrigo Moreno4
1Department of Biomedical Engineering and Health Systems, KTH Royal Institute of Technology, Hälsovägen 11C, SE-14157, Huddinge, Sweden. benklint@gmail.com.
Background:
As bone microstructure is known to impact bone strength, the aim of this in vitro study was to evaluate if the emerging photon-counting detector computed tomography (PCD-CT) technique may be used for measurements of trabecular bone structures like thickness, separation, nodes, spacing and bone volume fraction.
Methods:
Fourteen cubic sections of human radius were scanned with two multislice CT devices, one PCD-CT and one energy-integrating detector CT (EID-CT), using micro-CT as a reference standard. The protocols for PCD-CT and EID-CT were those recommended for inner- and middle-ear structures, although at higher mAs values: PCD-CT at 450 mAs and EID-CT at 600 (dose equivalent to PCD-CT) and 1000 mAs. Average measurements of the five bone parameters as well as dispersion measurements of thickness, separation and spacing were calculated using a three-dimensional automated region growing (ARG) algorithm. Spearman correlations with micro-CT were computed.
Results:
Correlations with micro-CT, for PCD-CT and EID-CT, ranged from 0.64 to 0.98 for all parameters except for dispersion of thickness, which did not show a significant correlation (p = 0.078 to 0.892). PCD-CT had seven of the eight parameters with correlations ρ > 0.7 and three ρ > 0.9. The dose-equivalent EID-CT instead had four parameters with correlations ρ > 0.7 and only one ρ > 0.9.
Conclusions:
In this in vitro study of radius specimens, strong correlations were found between trabecular bone structure parameters computed from PCD-CT data when compared to micro-CT. This suggests that PCD-CT might be useful for analysing bone microstructure in the peripheral human skeleton.