Related Experiment Video
Updated: Sep 3, 2025

Using Extraordinary Optical Transmission to Quantify Cardiac Biomarkers in Human Serum
Published on: December 13, 2017
A Polypyrrole/Nanoclay Hybrid Film for Ultra-Sensitive Cardiac Troponin T Electrochemical Immunosensor
Vicente P A Landim1, Marcos V Foguel1, Cecília M Prado1
1Biomedical Engineering Laboratory, Department of Biomedical Engineering, Federal University of Pernambuco, Av. Prof. Moraes Rego, 1235, Recife 50670-901, Brazil.
Abstract:
An electrochemical immunosensor based on a nanohybrid film of carboxylated polypyrrole and amine nanoclay was developed for label-free detection of the human cardiac troponin T (cTnT). The nanohybrid film was formed in situ on the surface of the glassy carbon electrode, followed by the covalent immobilization of anti-troponin T antibodies by glutaraldehyde. Morphological and chemical characterizations of the nanohybrid film were performed by scanning electron microscopy and Fourier-transform infrared spectroscopy. Under the optimized conditions, a calibration curve for cTnT in spiked serum was obtained by square wave voltammetry, and a low limit of detection and quantification was achieved (0.35 and 1.05 pg mL-1, respectively). This was the first time that this type of nanohybrid film was used in the development of an immunosensor for cTnT that proved to be a simple and efficient strategy for the manufacture of a label-free electrochemical device that could be applied in the diagnosis of acute myocardial infarction.