Using High-Resolution Ultrasound to Assess Post-Facial Paralysis Synkinesis-Machine Settings and Technical Aspects
Andreas Kehrer1, Marc Ruewe1, Natascha Platz Batista da Silva2
1Department of Plastic, Hand and Reconstructive Surgery, University Hospital Regensburg, 93053 Regensburg, Germany.
Background:
Synkinesis of the facial musculature is a detrimental sequalae in post-paralytic facial palsy (PPFP) patients. Detailed knowledge on the technical requirements and device properties in a high-resolution ultrasound (HRUS) examination is mandatory for a reliable facial muscle assessment in PPFP patients. We therefore aimed to outline the key steps in a HRUS examination and extract an optimized workflow schema.
Methods:
From December 2020 to April 2021, 20 patients with unilateral synkinesis underwent HRUS. All HRUS examinations were performed by the first author using US devices with linear multifrequency transducers of 4-18 MHz, including a LOGIQ E9 and a LOGIQ S7 XDclear (GE Healthcare; Milwaukee, WI, USA), as well as Philips Affinity 50G (Philips Health Systems; Eindhoven, the Netherlands).
Results:
Higher-frequency and multifrequency linear probes ≥15 MHz provided superior imaging qualities. The selection of the preset program Small Parts, Breast or Thyroid was linked with a more detailed contrast of the imaging morphology of facial tissue layers. Frequency (Frq) = 15 MHz, Gain (Gn) = 25-35 db, Depth (D) = 1-1.5 cm, and Focus (F) = 0.5 cm enhanced the image quality and assessability.
Conclusions:
An optimized HRUS examination protocol for quantitative and qualitative facial muscle assessments was proposed.