Comparison of PMT-based TF64 and SiPM-based Vereos PET/CT systems for 90 Y imaging and dosimetry optimization: A
Nicola Trotta1, Benoît Collette1, Céline Mathey1
1Department, of Nuclear Medicine, Hôpital Erasme, Université libre de Bruxelles (ULB), Brussels, Belgium.
Background:
Selective internal radiotherapy based on transarterial radioembolization (TARE) with yttrium-90 (90 Y) microspheres is an established treatment for primary or metastatic liver disease.
Purpose:
The objective of this work is to optimize the dosimetry of patients treated with 90 Y TARE, using positron emission tomography (PET) images.
Methods:
The NEMA 2012 PET phantom was filled with nearly 3.9 GBq of 90 Y activity and acquired at days 0, 3, 5, 7, and 9 on a classic time-of-flight PET/computed tomography (CT) scanner (Philips TF64) and on a silicon photomultiplier (SiPM)-based PET/CT scanner (Philips Vereos). Acquisitions were carried on following the guidelines proposed in a previously published multicentric trial and images were reconstructed by varying and combining the available parameters. Comparisons were performed to identify the best set(s) of parameters leading to the most accurate 90 Y-PET image(s), in terms of activity distribution. Then, for both scanners, the best images were analyzed with Simplicit90 Y, a personalized dosimetry software using multicompartmental Medical Internal Radiation Dose model. The comparison between measured and true doses allowed to identify the image granting the most consistent dose estimations and, therefore, to designate the set of parameters to be applied on patients' data for the reconstruction of optimized clinical images. Posttreatment dosimetry of four patients was then realized with Simplicit90 Y using optimized imaging datasets.
Results:
Based on activity distribution comparisons and dose estimations over phantom and patients data, the SiPM-based PET/CT system appeared more suitable than the photomultiplier tube-based TF64 for 90 Y-PET imaging. With the SiPM-based PET/CT system, reconstructed images with a 2-mm voxel size combined with the application of the point spread function correction led to the most accurate results for quantitative 90 Y measures.
Conclusions:
For the SiPM-based PET/CT scanner, an optimized set of reconstruction parameters has been identified and applied on patients' data in order to generate the most accurate image to be used for an improved personalized 90 Y-PET dosimetry, ensuring a reliable evaluation of the delivered doses.
More Related Videos
07:54Implantation and Monitoring by PET/CT of an Orthotopic Model of Human Pleural Mesothelioma in Athymic Mice
Published on: December 21, 2019
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies III: Computed Tomography