Related Experiment Video
Updated: Sep 3, 2025

Testing the Efficacy of Pharmacological Agents in a Pericardial Target Delivery Model in the Swine
Published on: July 7, 2016
Two cases of hemodynamic improvement by modulation of atrioventricular delay in cardiac operations
Akiko Tomita-Kobayashi1,2, Tomoko Fujimoto3, Shoko Takada3
1Anesthesiology Service, Sakurabashi-Watanabe Hospital, Osaka, Japan. akko_tomi1219@yahoo.co.jp.
Background:
Symptomatic sick sinus syndrome is one of the indications for pacemaker implantation, and we have to consider to program the pacemaker to an asynchronous pacing mode during an operation.
Case Presentation:
We reported two cases with a pacemaker implanted for sick sinus syndrome undergoing cardiac operation. We changed programming of the pacemaker to an asynchronous pacing mode (DOO) and modulated the programmed atrioventricular delay to avoid ventricular pacing, resulting in better hemodynamic condition. Although we observed premature ventricular contraction, no lethal arrhythmias induced by the R-on-T phenomenon were noted.
Conclusion:
Programming of the pacemaker to an asynchronous pacing mode and modulation of the programmed atrioventricular delay to avoid ventricular pacing may be an option for pacemaker management during an operation.
More Related Videos
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
08:42Cox-Maze IV Procedure Concomitant with Valvular Surgery In Situs Inversus Dextrocardia: A Single-Center Experience in China
Published on: February 11, 2022
Related Concept Videos
Cardiomyopathy VII: Pre and Post Operative Nursing Management
Aortic Regurgitation III: Medical Management
Heart Failure Drugs: Inhibitors of Renin-Angiotensin System
Dysrhythmias VI: Management of Dysrhythmias
Heart Failure V: Medical Management
Cardiomyopathy II: Dilated Cardiomyopathy