Improvement of an Automated Sample Injection System for Pillar Array Columns to Increase Analytical Reproducibility
Hiroshi Kuroki1, Hirotaka Koyama2, Makoto Tsunoda1
1Graduate School of Pharmaceutical Sciences, University of Tokyo, Tokyo 113-0033, Japan.
Abstract:
In our previous study, we developed an automatic sample injection system for pillar array columns for quantitative analysis. An autosampler was used to maintain a constant sample injection volume. However, the sample was diluted during injection using the autosampler, thus deteriorating the analytical reproducibility. In this study, we have substituted the autosampler with a syringe pump to overcome the abovementioned problem and improve the system. Sample dilution was avoided by filling the entire capillary with the sample at a constant rate. This improved system also increased the analytical reproducibility. In the previous system, the relative standard deviation (RSD) exceeded 17% of the peak height for coumarin dyes. In contrast, the improved system decreased the RSD to the range 1.2-1.8%. The analytical reproducibility was evaluated by using five types of amino acids. The RSD of each peak height was within 3.0%, confirming good reproducibility. These results indicate that the sample injection method developed in this study can be applied to biological sample analyses as a simple quantitative analysis method for pillar array columns.
Related Concept Videos
Gas Chromatography: Sample Injection Systems
Two primary injection methods are used...
High-Performance Liquid Chromatography: Elution Process