Angle Kappa agreement between Scheimpflug tomography, combined placido Scheimpflug and combined slit scanning placido
Mo'mena Ahmad A Awad-Allah1, Hesham Mohamed Gharieb1, Rania Gamal Eldin Zaki2
1Ain Shams University, Cairo, 11727, Egypt.
Purpose:
To compare the measured or calculated angle Kappa using Oculus pentacam HR, Sirius and Orbscan III devices.
Patients And Methods:
A prospective randomized cohort study, conducted on 47 eyes of 47 healthy orthotropic individuals, with an age range of 18-50 years and a corrected Snellen's distance visual acuity (CDVA) of 0.8 decimal or better. Angle Kappa is assessed directly using Orbscan® III software version 1.8.165.1. (Bausch and Lomb Rochester, New York, United States), while Pentacam® HR 1.21r.65 (Oculus Optikgeräte GmbH, Wetzlar, Germany) and Sirius device (CSO, version 3.2.1.60, Costruzione Strumenti Oftalmici, Florence, Italy) were used to calculate angle kappa indirectly.
Results:
Least mean difference of estimated angle Kappa was between Orbscan and Pentacam devices (- 0.18° ± 1.8), and it was statistically insignificant (p value = 0.1294). Differences between both Orbscan and Sirius, and Pentacam and Sirius were statistically significant (p value = 0.0004 and < 0.0001 consecutively). Bland Altman analysis showed a 95% confidence interval between Orbscan III and Pentacam of - 3.76 to 3.4 and between Orbscan III and Sirius of - 3.79 to 2.26.
Conclusion:
Pentacam parameters can be used as a reliable method to calculate angle kappa indirectly, without usage of any additional measurements from other machine. Sirius device parameters could also be used, but with less accurate results. A simple modification to those devices' software to calculate it, and incorporate it in the printout is possible, and highly recommended.
More Related Videos
10:25Author Spotlight: Integrating High-Resolution Intravital Imaging and MRI to Enhance Stereotactic Body Radiation Therapy Planning
Published on: April 12, 2024
10:17Switchable Acoustic and Optical Resolution Photoacoustic Microscopy for In Vivo Small-animal Blood Vasculature Imaging
Published on: June 26, 2017
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography