Verification of the effect of acquisition time for SwiftScan on quantitative bone single-photon emission computed
Takuro Shiiba1, Yuya Sekikawa2, Shinji Tateoka3
1Department of Molecular Imaging, School of Medical Sciences, Fujita Health University, 1-98, Dengakubo, Kutsukake-cho, Aichi, 470-1192, Toyoake, Japan. takuro.shiiba@fujita-hu.ac.jp.
Background:
SwiftScan single-photon emission computed tomography (SPECT) is a recently released scanning technique with data acquired when the detector is stationary and when it moves from one view to the next. The influence of scan time for using SwiftScan on quantitative bone SPECT remains unclear. This study aimed to clarify the effect of the scan time for SwiftScan SPECT on the image quality and quantification of bone SPECT compared to step and shoot mode (SSM) using 99mTc-filled anthropomorphic phantom (SIM2 bone phantom).
Materials And Methods:
Phantom SPECT/computed tomography (CT) images were acquired using Discovery NM/CT 860 (GE Healthcare) with a low-energy high-resolution sensitivity collimator. We used the fixed parameters (subsets 10 and iterations 5) for reconstruction. The coefficient of variation (CV), contrast-to-noise ratio (CNR), full width at half maximum (FWHM), and quantitative value of SwiftScan SPECT and SSM were compared at various acquisition times (5, 7, 17, and 32 min).
Results:
In the short-time scan (< 7 min), the CV and CNR of SwiftScan SPECT were better than those of SSM, whereas in the longtime scan (> 17 min), the CV and CNR of SwiftScan SPECT were similar to those of SSM. The FWHMs for SwiftScan SPECT (13.6-14.8 mm) and SSM (13.5-14.4 mm) were similar. The mean absolute errors of quantitative values at 5, 7, 17, and 32 min were 38.8, 38.4, 48.8, and 48.1, respectively, for SwiftScan SPECT and 41.8, 40.8%, 47.2, and 49.8, respectively, for SSM.
Conclusions:
SwiftScan on quantitative bone SPECT provides improved image quality in the short-time scan with quantification similar to or better than SSM. Therefore, in clinical settings, using SwiftScan SPECT instead of the SSM scan protocol in the short-time scan might provide higher-quality diagnostic images than SSM. Our results could provide vital information on the use of SwiftScan SPECT.