The probe-format graphite calorimeter, Aerrow, for absolute dosimetry in ultrahigh pulse dose rate electron beams
Alexandra Bourgouin1, Federico Keszti2, Andreas A Schönfeld3
1Dosimetry for Radiation Therapy and Diagnostic Radiology, Physikalisch-Technische Bundesanstalt (PTB), Braunschweig, Lower Saxony, Germany.
Purpose:
The purpose of this investigation is to evaluate the use of a probe-format graphite calorimeter, Aerrow, as an absolute and relative dosimeter of high-energy pulse dose rate (UHPDR) electron beams for in-water reference and depth-dose-type measurements, respectively.
Methods:
In this paper, the calorimeter system is used to investigate the potential influence of dose per pulses delivered up to 5.6 Gy, the number of pulses delivered per measurement, and its potential for relative measurement (depth-dose curve measurement). The calorimeter system is directly compared against an Advanced Markus ion chamber. The finite element method was used to calculate heat transfer corrections along the percentage depth dose of a 20-MeV electron beam. Monte Carlo-calculated dose conversion factors necessary to calculate absorbed dose-to-water at a point from the measured dose-to-graphite are also presented.
Results:
The comparison of Aerrow against a fully calibrated Advanced Markus chamber, corrected for the saturation effect, has shown consistent results in terms of dose-to-water determination. The measured reference depth is within 0.5 mm from the expected value from Monte Carlo simulation. The relative standard uncertainty estimated for Aerrow was 1.06%, which is larger compared to alanine dosimetry (McEwen et al. https://doi.org/10.1088/0026-1394/52/2/272) but has the advantage of being a real-time detector.
Conclusion:
In this investigation, it was demonstrated that the Aerrow probe-type graphite calorimeter can be used for relative and absolute dosimetries in water in an UHPDR electron beam. To the author's knowledge, this is the first reported use of an absorbed dose calorimeter for an in-water percentage depth-dose curve measurement. The use of the Aerrow in quasi-adiabatic mode has greatly simplified the signal readout, compared to isothermal mode, as the resistance was directly measured with a high-stability digital multimeter.
More Related Videos
14:11Quantification of Hydrogen Concentrations in Surface and Interface Layers and Bulk Materials through Depth Profiling with Nuclear Reaction Analysis
Published on: March 29, 2016
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
Related Concept Videos
Atomic Emission Spectroscopy: Overview
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....
Atomic Absorption Spectroscopy: Instrumentation
The atomizer used in AAS can be either a flame atomizer or an...
Atomic Emission Spectroscopy: Lab
Atomic Emission Spectroscopy: Instrumentation
Atomic Absorption Spectroscopy: Radiation and Light Sources
Two common narrow-range 'line' sources used in AAS are hollow-cathode lamps (HCLs) and...