Comparison of the CytoSorb® 300 mL and Jafron HA380 hemoadsorption devices: an in vitro study
Axel Nierhaus1,2, Jesus Morales3, Daniel Wendt3,4
1Clinic for Anesthesiology and Intensive Care Medicine, University Medical Center Hamburg, Hamburg, Germany.
Introduction:
We performed an analysis of two blood purification systems to determine their performance for removing interleukins (ILs)-6 and 10, tumor necrosis factor (TNF)-α and monocyte chemoattractant protein (MCP)-1 from blood.
Material And Methods:
An in vitro hemoperfusion blood recirculation circuit was used to compare the CytoSorb® 300 mL (CytoSorbents Inc., Princeton, NJ) and Jafron HA 380 (Jafron Biomedical Co., Ltd., Zhuhai City, China) devices. The removal of purified recombinant human IL-6, IL-10, TNFα and MCP-1 by the adsorbers was compared at various timepoints. Three runs were completed and removal was evaluated as the mean area under the curve (AUC).
Results:
Both devices showed effective removal of the tested cytokines. IL-6, IL-10, TNFα and MCP-1 were removed faster and to a higher extent by the CytoSorb® 300 mL device. At maximal time of 12 h, overall removal according to AUC of remaining concentrations was significantly lower with CytoSorb® 300 mL compared with HA 380 (IL-6: 1075.5 ± 665.9 vs. 4345.1 ± 1499.3 (p = 0.01), IL-10: 5065.7 ± 882.5 vs. 11,939.7 ± 4523.1 (p = 0.03), TNF-α: 6519.9 ± 997.6 vs. 10,303.7 ± 2347.0 (p = 0.03) and MCP-1: 278.9 ± 40.7 vs. 607.3 ± 84.4 (p = 0.001)).
Conclusions:
Both the CytoSorb® and the Jafron HA 380 devices are capable of removing cytokines from blood in a benchtop model. The CytoSorb® 300 device was significantly more efficient achieving the bulk of the removal in the first 120 min.
More Related Videos
Related Concept Videos
Dialysis
Renal Drug Clearance: Comparison Between Renal Excretion Methods
Renal clearance is often associated with the renal glomerular filtration rate (GFR), which represents the rate at which plasma is filtered through the glomeruli in the kidney. When drug reabsorption is minimal and there is no active secretion, renal clearance is closely related to the...
Methods for Studying Drug Absorption: In vitro
The diffusion cell method uses a two-compartment cell, including a donor compartment with the drug solution, which simulates the environment where the drug is applied, and a receptor compartment with a buffer solution, which simulates the environment...