Clinical assessment of image quality, usability and patient comfort in dedicated spiral breast computed tomography
Cynthia Sabrina Schmidt1, Caroline Zellweger1, Jann Wieler1
1Institute of Diagnostic and Interventional Radiology, University Hospital Zurich, Raemistrasse 100, CH-8091 Zurich, Switzerland.
Objective:
To investigate aspects of image quality, feasibility and patient comfort in dedicated spiral breast computed tomography (B-CT) in a large patient cohort.
Methods:
This retrospective study was approved by the institutional review board. 2418 B-CT scans from 1222 women examined between 04/16/2019 and 04/13/2022 were analyzed. Patients evaluated their comfort during the examination, radiographers carrying out the scans evaluated the patient's mobility and usability of the B-CT device, whereas radiologists assessed lesion contrast, detectability of calcifications, breast coverage and overall image quality. For semi-quantitative assessment, a Likert-Scale was used and statistical significance and correlations were calculated using ANOVAs and Spearman tests.
Results:
Comfort, mobility and usability of the B-CT were rated each with either "no" or "negligible" complaints in >99%. Image quality was rated with "no" or "negligible complaints" in 96.7%. Lesion contrast and detectability of calcifications were rated either "optimal" or "good" in 92.6% and 98.4%. "Complete" and "almost complete" breast coverage were reported in 41.9%, while the pectoral muscle was found not to be covered in 56.0%. Major parts of the breast were not covered in 2.1%. Some variables were significantly correlated, such as age with comfort (ρ = -0.168, p < .001) and mobility (ρ = -0.172, p < .001) as well as patient weight with lesion contrast (ρ = 0.172, p < .001) and breast coverage (ρ = -0.109, p < .001).
Conclusions:
B-CT provides high image quality and contrast of soft tissue lesions as well as calcifications, while covering the pre-pectoral areas of the breast remains challenging. B-CT is easy to operate for the radiographer and comfortable for the majority of women.
More Related Videos
15:48Tracking the Mammary Architectural Features and Detecting Breast Cancer with Magnetic Resonance Diffusion Tensor Imaging
Published on: December 15, 2014
06:50Troubleshooting FoCUS Image Acquisition: Patient Positioning, Transducer Manipulation, and Image Optimization
Published on: March 3, 2023
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System V: CT
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET