Comparative analysis of patient scatter to phantom scatter for computed tomography systems
1Biomedical Technology Services, Princess Alexandra Hospital, Ipswich Rd, Woolloongabba, QLD, 4102, Australia. stephen.edwards@health.qld.gov.au.
Abstract:
Current computed tomography (CT) shielding practices are largely based on calculations of scattered radiation emitted from an acrylic head or body phantom, such as the Computed Tomography Dose Index (CTDI) phantom, or anthropomorphic phantoms of these two anatomical categories. This report considers the difference in scattered air kerma or dose from phantom models, to actual patient scatter under a variety of clinical scan conditions. Empirical patient scatter measurements recorded at different positions around the gantry, for 3 different CT scanners, resulted in average patient scatter fractions per unit Dose Length Product (DLP: mGy cm) at 1 m from the isocentre of 0.09 ± 0.03 and 0.17 ± 0.04 µGy (mGy cm)-1 for head and body scans respectively. For the purposes of shielding design and scattered dose estimates to staff it is recommended that a single standard deviation be applied to these averages in the continued interest of conservatism. These values are reasonably comparative to the widely published scatter fractions by the National Council of Radiological Protection using the CTDI phantom, and the British Institute of Radiology using head and body anthropomorphic phantoms.
More Related Videos
08:41Patient-Specific Polyvinyl Alcohol Phantom Fabrication with Ultrasound and X-Ray Contrast for Brain Tumor Surgery Planning
Published on: July 14, 2020
07:13Author Spotlight: An Efficient and Robust Software for Automated Fusion of Multiple Preclinical Imaging Modalities
Published on: October 27, 2023
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...