Related Experiment Video
Updated: Jul 21, 2026

Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies
Published on: February 6, 2019
Use of a PLA sleeve to remove electron enhancement in superficial X-ray therapy
1Rad-e-Cal Ltd, Melbourne, Australia. c.barbagallo@alfred.org.au.
Abstract:
A newly installed superficial X-ray unit was found to produce enhanced electron dose at the skin surface. The ACPSEM kilovoltage dosimetry recommendations suggest using nail varnish within the treatment cones as a method to reduce this dose. In this study, a 3D PLA sleeve was produced and used as an alternative to the nail varnish for energies between 55 and 100 kV. Further, plastic wrap was also investigated as an alternative method to reduce dose. It was found that a 1 mm printed sleeve, inserted into the treatment cone sufficiently reduced the enhanced dose as measured with a thin-window Exradin chamber to within 3.3% of the dose measured with a Farmer-type ionisation chamber. The use of plastic wrap also reduced the enhanced dose, but impracticalities in its use make it non-viable for routine clinical use.
More Related Videos
07:48High Spatial Resolution Chemical Imaging of Implant-Associated Infections with X-ray Excited Luminescence Chemical Imaging Through Tissue
Published on: September 30, 2022
13:48Reduction of Radiation Exposure during Endovascular Treatment of Peripheral Arterial Disease Combining Fiber Optic RealShape Technology and Intravascular Ultrasound
Published on: April 21, 2023
Related Concept Videos
Preparation of Samples for Electron Microscopy
X-ray Imaging
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...