Related Experiment Video
Updated: Sep 2, 2025

Author Spotlight: 3D Movement Assessment of Maxillary Posterior Teeth in Clear Aligner Treatment
Published on: February 23, 2024
The "kickstand rod" technique for correction of coronal malalignment: two-year clinical and radiographic outcomes
Varun Puvanesarajah1, Micheal Raad2, Fthimnir M Hassan3
1The Daniel and Jane Och Spine Hospital, New York Presbyterian, Columbia University Medical Center, 5141 Broadway, NY, NY, 10034, USA. vpuvanesarajah@gmail.com.
Purpose:
Restoring coronal alignment in spine deformity patients has been shown to play an important role in improving patient reported outcomes (PRO). Recently, the "kickstand rod" (KSR) technique was developed as a novel coronal correction method in complex spine deformity cases. The goal of the present study was to assess outcomes of this technique at two years of follow-up.
Methods:
Consecutive, unique adult patients who underwent KSR constructs for coronal spinal malalignment between 2015 and 2019 with a minimum 2 year clinical and radiographic follow-up were identified. A KSR construct includes a more laterally placed iliac screw and additional rod that effectively depresses the ipsilateral ilium/pelvis for coronal correction, while serving as a buttress to prevent future loss of correction. Outcomes included revision for instrumentation-related complications, radiographic alignment, and PROs.
Results:
Twenty patients were included with a mean age of 54 years [range: 20-73 years]. Mean follow-up time was 2.5 years [range: 2.0-5.0]. Mean number of levels fused was 17.3 [range: 10-24]. There were significant improvements in coronal alignment (CVA: 5.8 cm ± 2.6 cm vs. 1.7 cm ± 1.5 cm), sagittal alignment (SVA: 5.6 cm ± 5.9 cm vs. 1.6 cm ± 2.5 cm) and major Cobb angle (55º ± 32 vs. 26º ± 21) maintained at 2 years (p < 0.05). One patient experienced an asymptomatic fracture at the shank of the KSR iliac screw. There were significant improvements in Oswestry Disability Index and SRS-22 domains (p < 0.05).
Conclusion:
The KSR technique is a safe and effective method for correcting coronal malalignment in complex spinal deformity patients with no revisions specific for the KSR or iliac screw and significantly improved PROs at a minimum two-year follow-up.