Evaluation of a New Smartphone Powered Low-cost Pulse Oximeter Device
Hundessa Daba Nemomssa1, Hakkins Raj1
1School of Biomedical Engineering, Jimma Institute of Technology, Jimma University, Jimma, Oromia, Ethiopia.
Background:
Measurement of blood oxygen saturation is a vital part of monitoring coronavirus 2019 (COVID-19) patients. Pulse oximetry is commonly used to measure blood oxygen saturation and pulse rate for appropriate clinical intervention. But the majority of direct-to-consumer grade pulse oximeters did not pass through in-vivo testing, which results in their accuracy being questionable. Besides this, the ongoing COVID-19 pandemic exposed the limitations of the device in resource limited areas since independent monitoring is needed for COVID-19 patients. The purpose of this study was to perform an in-vivo evaluation of a newly developed smartphone powered low-cost pulse oximeter.
Methods:
The new prototype of a smartphone powered pulse oximeter was evaluated against the standard pulse oximeter by taking measurements from fifteen healthy volunteers. The accuracy of measurement was evaluated by calculating the percentage error and standard deviation. A repeatability and reproducibility test were carried out using the ANOVA method.
Results:
The average accuracy for measuring spot oxygen saturation (SPO2) and pulse rate (PR) was 99.18% with a standard deviation of 0.57 and 98.78% with a standard deviation of 0.61, respectively, when compared with the standard pulse oximeter device. The repeatability and reproducibility of SPO2 measurements were 0.28 and 0.86, respectively, which is in the acceptable range.
Conclusion:
The new prototype of smartphone powered pulse oximeter demonstrated better performance compared to the existing low-cost fingertip pulse oximeters. The device could be used for independent monitoring of COVID-19 patients at health institutions and also for home care.
More Related Videos
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
12:51Evaluation of Commercial-Off-The-Shelf Wrist Wearables to Estimate Stress on Students
Published on: June 16, 2018
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation