Validation of Virtual Monochromatic Images and Effect of Body Size Obtained Using a Rapid kVp-switching Dual-energy
Tung-Hsin Wu1, Yun-Lung Ting, Yi-Shuan Hwang2
1Service Unit, Department of Biomedical Imaging and Radiological Sciences, National Yang Ming Chiao Tung University, Taipei, Taiwan, R.O.C.
Abstract:
The objective of this paper is to validate virtual monochromatic computed tomography (CT) numbers and the effect of the body size of insert materials in phantoms on the findings of a dual-energy CT scanner. The material inserted in the phantom simulates human organs. This study investigated the effect of different body sizes on CT numbers to understand the accuracy of dual-energy CT. The effect of body size on virtual monochromatic CT numbers was investigated using a QRM phantom. The true monochromatic CT numbers of insert materials were calculated from coefficients obtained using NIST XCOM. The true Z eff values were supplied by phantom manufacturers or computed using Mayneord's equation. The virtual monochromatic CT numbers of insert materials in both the phantoms varied with energy. The CT numbers of materials with a Z eff of >7.42 (water Z eff ) and <7.42 decreased and increased with energy, respectively. The CT numbers were affected by phantom size as a function of energy. For water, tissues, and air, the CT numbers in the XL phantom were considerably larger than those in other phantom sizes at 40 keV. Body size affected the CT numbers, particularly for the XL size and at low energies. For all materials, the magnitude of difference between the measured and true CT numbers was related to the Z eff of the materials, potentially because the photoelectric effect is more prominent at low energies for materials with a higher Z eff . The difference in CT numbers appeared to be dependent on position. The true and measured Z eff agreed to within 6% for all the materials except the SR2 brain, for which the discrepancy was 25%.
More Related Videos
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
X-ray Imaging