Validating the accuracy of multispectral metal artifact suppressed diffusion-weighted imaging
John P Neri1, Matthew F Koff1, Kevin M Koch2
1MRI Research Laboratory, Hospital for Special Surgery, New York, New York, USA.
Medical Physics
|August 11, 2022
Summary
Diffusion-weighted imaging with multi-acquisition with variable resonance image combination (DWI-MAVRIC) shows good accuracy for total hip arthroplasty imaging. While linearity is slightly reduced compared to conventional methods, it offers reliable results for assessing microstructural changes.
Area of Science:
- Musculoskeletal MRI
- Medical Imaging Physics
Background:
- Diffusion-weighted imaging (DWI) quantifies water diffusion (ADC) for microstructural analysis in tissues.
- Conventional DWI in total hip arthroplasty (THA) is challenged by metallic artifacts from echo-planar imaging (EPI).
- Multispectral imaging (MSI) techniques like MAVRIC reduce artifacts, and PROPELLER FSE with DWI-MAVRIC aims to further mitigate distortion and dropout.
Purpose of the Study:
- To assess the quantitative accuracy of DWI-MAVRIC compared to standard diffusion sequences.
- To evaluate the repeatability and reliability of DWI-MAVRIC in musculoskeletal imaging applications.
Main Methods:
- Diffusion phantom scans using DWI-EPI and DWI-MAVRIC on two 1.5T MRI scanners.
- Acquisitions performed in axial and coronal planes with varying spatial offsets to assess positioning effects.
- Intraday, interday, and inter-scanner repeatability evaluated; gradient nonlinearity correction (GNC) impact assessed.
Main Results:
- DWI-MAVRIC demonstrated excellent correlation (r=1.00) but slightly reduced linearity (slope=0.91-0.94) versus DWI-EPI (slope=0.99-1.01).
- Minor ADC bias (>5%) observed in DWI-MAVRIC at lowest ADC values; no variation with DWI-MAVRIC parameters.
- DWI-MAVRIC showed inferior repeatability (intraday: 13.3, interday: 44.7, inter-scanner: 110 μm²/s) compared to DWI-EPI (3.18, 19.2, 20.2 μm²/s).
Conclusions:
- DWI-MAVRIC offers good linearity and acceptable intra- and interday repeatability for musculoskeletal applications.
- The technique shows promise for artifact reduction in THA imaging, despite lower repeatability than conventional DWI.