Related Experiment Video
Updated: Sep 1, 2025

Monitoring Lung Function with Electrical Impedance Tomography in the Intensive Care Unit
Published on: September 6, 2024
Use of a Portable Electronic Interface Improves Clinical Handoffs and Adherence to Lung Protective Ventilation
Neil R Euliano1, Paul Stephan1, Konstantinos Michalopoulos1
1Convergent Engineering, Inc, Gainesville, FL, USA.
Background:
Mechanical ventilation (MV) is used to support patients with respiratory impairment. Evidence supports the use of lung-protective ventilation (LPV) during MV to improve outcomes. However, studies have demonstrated poor adherence to LPV guidelines. We hypothesized that an electronic platform adapted to a hand-held tablet receiving real-time ventilatory parameters could increase clinician awareness of key LPV parameters. Furthermore, we speculated that an electronic shift-change tool could improve the quality of clinician handoffs.
Methods:
Using a specially designed Wi-Fi dongle to transmit data from three ventilators and a respiratory monitor, we implemented a system that displays data from all ventilators under the care of a Respiratory Care Practitioner (RCP) on an electronic tablet. In addition, the tablet created a handoff checklist to improve shift-change communication. In a simulated ICU environment, we monitored the performance of eight RCPs at baseline and while using the system.
Results:
Using the system, the time above guideline Pplat decreased by 74% from control, and the time outside the VT range decreased by 60% from control, p = 0.007 and 0.015, respectively. The handoff scores improved quality significantly from 2.8 to 1.6 on a scale of 1 to 5 (1 being best), p = 0.03.
Conclusion:
In a simulated environment, an electronic RT tool can significantly improve shift-change communication and increase the RCP's level of LPV adherence.
More Related Videos
Related Concept Videos
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Types of Reports I: Hands-off Report
Following are the key components and categories of hand-off reports:
Purpose and Process:
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
PPE Use in Healthcare Settings I: Donning
Cardiopulmonary Resuscitation II: ACLS Airway Management
PPE Use in Healthcare Settings II: Doffing