Evaluation of preanalytical and postanalytical phases in clinical biochemistry laboratory according to IFCC
Nergiz Zorbozan1, Orçun Zorbozan2
1Kemalpaşa State Hospital, Medical Biochemistry, İzmir, Turkey.
Introduction:
The aim of the study was to determine the current state of laboratory's extra-analytical phase performance by calculating preanalytical and postanalytical phase quality indicators (QIs) and sigma values and to compare obtained data according to desired quality specifications and sigma values reported by The International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) Working Group - Laboratory errors and Patient Safety.
Materials And Methods:
Preanalytical and postanalytical phase data were obtained through laboratory information system. Rejected samples in preanalytical phase were grouped according to reasons for rejection and frequencies were calculated both monthly and for 2019. Sigma values were calculated according to "short term sigma" table.
Results:
The number of rejected samples in laboratory was 643 out of 191,831 in 2019. Total preanalytical phase rejection frequency was 0.22%. According to the reasons for rejection, QIs and sigma values were: "Samples with excessive transportation time": 0.0036 and 5.47; "Samples collected in wrong container" 0.02 and 5.11. In December, QIs and sigma values were: "Samples with excessive transportation time": 0.01 and 5.34; "Samples collected in wrong container": 0.03 and 4.98. The postanalytical QIs and sigma values were: "Reports delivered outside the specified time": 0.34 and 4.21; "Turn around time of potassium": 56 minute and 3.84, respectively. There were no errors in "Critical values of inpatients and outpatients notified after a consensually agreed time".
Conclusions:
Extra-analytical phase was evaluated by comparing it with the latest quality specifications and sigma values which will contribute to improving the quality of laboratory medicine.
More Related Videos
11:17Multiplex Therapeutic Drug Monitoring by Isotope-dilution HPLC-MS/MS of Antibiotics in Critical Illnesses
Published on: August 30, 2018
14:45Enumeration of Major Peripheral Blood Leukocyte Populations for Multicenter Clinical Trials Using a Whole Blood Phenotyping Assay
Published on: September 16, 2012
Related Concept Videos
Contaminants and Errors
Another key consideration is determining the appropriate number of samples required to...
Systematic Error: Methodological and Sampling Errors
Sampling errors originate from improper sampling methods or the wrong sample population. These errors can be minimized by refining the sampling strategy. Defective instruments or faulty calibrations are the sources of instrumental...
Data Validation
Key parameters for method validation include:
Errors occurring during blood pressure monitoring
Several factors...
Quality Assurance
Preclinical Development: Overview