Is There Any Difference Between a Spherical Marker and a Simple Coin for Hip Replacement Digital Planning?
Giancarlo Cavalli Polesello1, Thiago Tronco Salerno1, Marcelo Cavalheiro de Queiroz1
1Departamento de Ortopedia e Traumatologia da Irmandade da Santa Casa de Misericórdia de São Paulo (ISCMSP), São Paulo, SP, Brasil.
Revista Brasileira De Ortopedia
|August 15, 2022
Summary
The coin marker is most accurate for hip arthroplasty planning, especially when placed between the thighs. Researchers recommend using a coin marker on the exam table for comparison due to its reliable measurements.
Area of Science:
- Orthopedic Surgery
- Radiology
- Medical Imaging
Background:
- Accurate preoperative planning is crucial for successful hip arthroplasty.
- Radiographic calibration relies on precise measurement tools.
- Metallic markers are used to ensure accuracy in imaging.
Purpose of the Study:
- To compare the accuracy of spherical and coin-shaped metallic markers for radiographic calibration.
- To evaluate marker performance in preoperative planning for hip arthroplasty.
- To determine the optimal marker type and placement for hip imaging.
Main Methods:
- 33 patients undergoing hip arthroplasty were included.
- Spherical and coin markers (25mm diameter) were placed at various anatomical locations.
- Image diameters were measured, and statistical analysis was performed.
Main Results:
- Coin markers placed between the thighs showed higher accuracy (50-60%) compared to spheres.
- Markers on the greater trochanter had variable visualization rates (up to 57.6% not visualized).
- Coin and sphere markers on the exam table showed minimal size reduction and easier positioning.
Conclusions:
- The coin marker positioned between the thighs is recommended for hip arthroplasty radiographic calibration.
- Using a coin marker on the exam table for comparison is suggested.
- A slight size reduction (8%) of markers should be considered in measurements.