Measuring the 50% Haemolytic Complement (CH50) Activity of Serum
Published on: March 29, 2010
A sensitive haemolytic assay of mouse complement
Under appropriate conditions the haemolytic activity of mouse complement can be measured in a 51Cr release assay. Modifications that increase the sensitivity are the following: (a) mouse 7 S anti-SRBC antibodies as amboceptor; (b) low amount of target cells; (c) ionic strength of 0.13 M NaCl in the test medium; (d) incubation temperature of 30 degrees C; (e) incubation time of 90 min; (f) total reaction volume of 1 ml.
Under appropriate conditions the haemolytic activity of mouse complement can be measured in a 51Cr release assay. Modifications that increase the sensitivity are the following: (a) mouse 7 S anti-SRBC antibodies as amboceptor; (b) low amount of target cells; (c) ionic strength of 0.13 M NaCl in the test medium; (d) incubation temperature of 30 degrees C; (e) incubation time of 90 min; (f) total reaction volume of 1 ml.