A comparative study of single-channel signal processing methods in fetal phonocardiography
Katerina Barnova1, Radana Kahankova1, Rene Jaros1
1Department of Cybernetics and Biomedical Engineering, Faculty of Electrical Engineering and Computer Science, VSB-Technical University of Ostrava, Ostrava, Czechia.
Abstract:
Fetal phonocardiography is a non-invasive, completely passive and low-cost method based on sensing acoustic signals from the maternal abdomen. However, different types of interference are sensed along with the desired fetal phonocardiography. This study focuses on the comparison of fetal phonocardiography filtering using eight algorithms: Savitzky-Golay filter, finite impulse response filter, adaptive wavelet transform, maximal overlap discrete wavelet transform, variational mode decomposition, empirical mode decomposition, ensemble empirical mode decomposition, and complete ensemble empirical mode decomposition with adaptive noise. The effectiveness of those methods was tested on four types of interference (maternal sounds, movement artifacts, Gaussian noise, and ambient noise) and eleven combinations of these disturbances. The dataset was created using two synthetic records r01 and r02, where the record r02 was loaded with higher levels of interference than the record r01. The evaluation was performed using the objective parameters such as accuracy of the detection of S1 and S2 sounds, signal-to-noise ratio improvement, and mean error of heart interval measurement. According to all parameters, the best results were achieved using the complete ensemble empirical mode decomposition with adaptive noise method with average values of accuracy = 91.53% in the detection of S1 and accuracy = 68.89% in the detection of S2. The average value of signal-to-noise ratio improvement achieved by complete ensemble empirical mode decomposition with adaptive noise method was 9.75 dB and the average value of the mean error of heart interval measurement was 3.27 ms.
Related Concept Videos
Fetal Circulation
Two umbilical arteries transport blood from the fetus to the placenta. At the placenta, the blood absorbs oxygen and nutrients while simultaneously eliminating waste products. This oxygen-enriched and nutrient-rich blood then returns to the fetus through one...
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for...
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion,...