Related Experiment Video
Updated: Aug 31, 2025

3D Imaging of Soft-Tissue Samples using an X-ray Specific Staining Method and Nanoscopic Computed Tomography
Published on: October 24, 2019
Near-lossless compression of PET/CT images using singular value decomposition
Jagrati Chaudhary1, Anil Kumar Pandey1, Param D Sharma2
1Department of Nuclear Medicine, All India Institute of Medical Sciences.
Aims And Objectives:
The aim of the study is to compare the single matrix approach and slice-by-slice approach for computing singular value decomposition (SVD) to achieve near-lossless compression of PET/CT images.
Materials And Methods:
The parameters used for comparison were SVD computation time, percentage compression and percentage difference between ROI counts on compressed and original images. SVD of 49 F-18-FDG PET/CT studies (33 370 PET/CT images) was computed using both approaches. The smaller singular values contributing insignificant information to the image were truncated, and then, the compressed image was reconstructed. A mask (101 × 101pixels) was used to extract the ROI counts from compressed and original images. Two nuclear medicine physicians compared compressed images with their corresponding original images for loss of clinical details and the presence of generated artifacts. Structural Similarity Index Measure, blur, brightness, contrast per pixel and global contrast factor were used for objective assessment of image quality. Wilcoxon test was applied to find a statistically significant difference between the parameters used for comparison at alpha = 0.05.
Results:
Nuclear medicine physicians found compressed image identical to the corresponding original image. The values of comparation parameters were significantly larger for the single matrix approach in comparison with the slice-by-slice approach. The maximum percentage error between the compressed image and original image was less than 5%.
Conclusions:
Up to 64 % and 44% near-lossless compression of PET and CT images were achieved, respectively using the slice-by-slice approach, and up to 58 and 53% near-lossless compression of PET and CT images were achieved respectively using the single matrix approach.
More Related Videos
07:53Author Spotlight: Advancing 3D Modeling for Enhanced Diagnosis and Treatment of Pulmonary Nodules in Early-Stage Lung Cancer
Published on: October 13, 2023
08:02Author Spotlight: Enhanced Quantification of Cardiovascular Calcification Progression for Longitudinal Micro PET/CT Studies in Small Research Animals
Published on: November 15, 2024
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT