Relapsing low-flow alarms due to suboptimal alignment of the left ventricular assist device inflow cannula
Casper F Zijderhand1,2, Wiebe G Knol1,3, Ricardo P J Budde3
1Department of Cardiothoracic Surgery, Thoraxcenter, Erasmus MC, University Medical Center Rotterdam, Rotterdam, Netherlands.
Objectives:
This retrospective study investigated the correlation between the angular position of the left ventricular assist device (LVAD) inflow cannula and relapsing low-flow alarms.
Methods:
Medical charts were reviewed of all patients with HeartMate 3 LVAD support for relapsing low-flow alarms. A standardized protocol was created to measure the angular position with a contrast-enhanced computed tomography scan. Statistics were done using a gamma frailty model with a constant rate function.
Results:
For this analysis, 48 LVAD-supported patients were included. The majority of the patients were male (79%) with a median age of 57 years and a median follow-up of 30 months (interquartile range: 19-41). Low-flow alarm(s) were experienced in 30 (63%) patients. Angulation towards the septal-lateral plane showed a significant increase in low-flow alarms over time with a constant rate function of 0.031 increase in low-flow alarms per month of follow-up per increasing degree of angulation (P = 0.048). When dividing this group using an optimal cut-off point, a significant increase in low-flow alarms was observed when the septal-lateral angulation was 28° or more (P = 0.001). Anterior-posterior and maximal inflow cannula angulation did not show a significant difference.
Conclusions:
This study showed an increasing number of low-flow alarms when the degrees of LVAD inflow cannula expand towards the septal-lateral plane. This emphasizes the importance of the LVAD inflow cannula angular position to prevent relapsing low-flow alarms with the risk of diminished quality of life and morbidity.
More Related Videos
03:40Point-of-Care Ultrasound for Peripheral Veno-Arterial Extracorporeal Membrane Oxygenation Without Left Ventricular Venting
Published on: January 17, 2025
08:49Author Spotlight: Enhancing Graft Viability Assessment Through Quantitative Metrics and Innovative Reservoir Systems
Published on: August 2, 2024
Related Concept Videos
Aortic Regurgitation I: Introduction
Aortic Regurgitation IV: Nursing Management
Errors occurring during blood pressure monitoring
Several factors...
Aortic Regurgitation III: Medical Management
Mitral Stenosis I: Introduction
Imbalances in Cardiac Output
CHF can occur due to the failure of either side of the heart. Left-side failure leads to pulmonary congestion—the right side continues to send...