An experimental study of flexural strength and hardness of zirconia and their relation to crown failure loads
Tariq F Alghazzawi1, Gregg M Janowski2, Alan W Eberhardt3
1Associate Professor, Department of Substitutive Dental Sciences, Taibah University, Madina, Saudi Arabia; Fullbright Scholar, Department of Materials Science and Engineering, The University of Alabama at Birmingham, Birmingham, USA.
Statement Of Problem:
Dental zirconia is often marketed and selected for restorative use based upon the biaxial flexural strength of prefabricated specimens (disks) without considering other mechanical and physical properties.
Purpose:
The purpose of this in vitro study was to test whether 4-point flexural strength, biaxial flexural strength, and/or hardness may correlate with failure loads for crowns made of different zirconia materials.
Material And Methods:
Three brands of zirconia (BruxZir, Cercon, and Katana) were used to fabricate anatomically contoured crowns, rectangular bars, and circular disks. The sample size was n=15 specimens per zirconia brand and specimen shape. The bars were tested for 4-point flexural strength by using the 4-point bending (4PB) test and Vickers hardness (VH), while the disks were tested for biaxial flexural strength by using a piston-on-3 ball (POB) test. Crowns were attached to resin abutments and compressed with a steel spherical indenter through a polyethylene sheet to assess the failure loads by using the "crunch the crown" (CTC) test. One-way ANOVA (α=.05) was used to test for statistically significant differences between groups, and Weibull analysis was used to assess the variability of the measured flexural strengths, failure load, and hardness.
Results:
Statistical differences (P<.001) were found in comparing Cercon, BruxZir, and Katana ([4260 N=4186 N]>3195 N, respectively) with the CTC test and (396 MPa>[281 MPa=275 MPa], respectively) the 4PB test. No statistical differences (P=.084) were found among the zirconia brands (Cercon [384 MPa]=Bruxzir [359 MPa]=Katana [416 MPa]) for the POB test. No significant correlations (P>.05) were found between the 4PB, POB, or VH tests and the corresponding CTC test. The Weibull modulus varied for the different specimen geometries.
Conclusions:
The piston-on-3 ball, 4-point bending, and Vickers hardness test results were not correlated with the corresponding crunch-the-crown test.
Related Concept Videos
Flexural Stress
Hooke's Law states that within the material's elastic limits, stress is directly proportional to strain. In a member experiencing a bending moment, the strain at any point is relative to...
Strength of Cement
For compressive strength tests, ASTM C 109-05 standards prescribe a cement-sand mix ratio of 1:2.75 and a water/cement ratio of 0.485 for making 2-inch cubes. These cubes are mixed, cast, and cured in saturated lime water at 23°C until testing. Flexural strength testing, outlined in...
Stresses under Combined Loadings
The process begins by slicing the tube at critical points and analyzing the internal forces and stress components at these sections, focusing on the centroid. Normal stresses, generated by axial forces and bending moments, are either compressive or tensile and vary across the section from...
Residual Stresses in Bending
Impact Strength of Concrete
Behavior of Concrete Under Compressive Load
As the concrete specimen fractures under...