A Novel Triple-Bladder Cuff Method for Blood Pressure Estimation Based on Pulse Wave Dynamics in Brachial Artery:
Objective:
The objective of this study was to develop a novel triple-bladder cuff method for accurate and automated estimation of systolic (SBP) and diastolic (DBP) blood pressure and validate its reliability in animal experiments.
Methods:
The cuff is composed of three bladders each measured one-third the width of a conventional BP cuff, which are designed to measure oscillatory pulsation at the proximal, middle, and distal segments of the upper arm. This structure allows evaluation of the pulse wave propagation in the brachial artery under the cuff. SBP is estimated (SBPe) by detecting resumption of systolic arterial flow based on statistical similarity in oscillatory pulse traces between the proximal and distal segments. DBP is estimated (DBPe) based on the relation between pulse wave velocity and transmural pressure at diastole in the brachial artery. In 7 anesthetized goats, we compared SBPe and DBPe to reference SBP and DBP, respectively, measured by an intra-arterial catheter. BP was perturbed by infusing nitroprusside or noradrenaline.
Results:
SBP correlated strongly with SBPe in each animal [mean coefficient of determination (R2) = 0.98 ± 0.01]. Mean ± standard deviation of errors between SBP and SBPe was 0.0 ± 4.9 mmHg. DBP correlated strongly with DBPe in each animal (R2 = 0.96 ± 0.03). Mean ± standard deviation of errors between DBP and DBPe was 0.0 ± 6.3 mmHg.
Conclusion:
This method estimates SBP and DBP with acceptable accuracy.
Significance:
Accurate and automated BP estimation by this method may potentially optimize antihypertensive treatment in patients with hypertension.
More Related Videos
14:09Tilt Testing with Combined Lower Body Negative Pressure: a "Gold Standard" for Measuring Orthostatic Tolerance
Published on: March 21, 2013
08:42Ultrasound Assessment of Endothelial-Dependent Flow-Mediated Vasodilation of the Brachial Artery in Clinical Research
Published on: October 22, 2014
Related Concept Videos
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Pre-Procedural Guidelines for Assessing Blood Pressure
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...