Adiabatic inversion radiofrequency (RF) pulses can be transformed into refocusing pulses. These modified pulses maintain stability against RF field variations and are useful for surface coil imaging.
Area of Science:
- Magnetic Resonance Imaging
- Radiofrequency Pulse Design
Background:
- Adiabatic inversion radiofrequency (RF) pulses are crucial for magnetic resonance imaging (MRI) experiments.
- These pulses are sensitive to variations in RF field strength, limiting their application in certain scenarios.
Purpose of the Study:
- To convert adiabatic 180-degree inversion RF pulses into 180-degree refocusing pulses.
- To develop RF pulses that are insensitive to large RF field variations and cover a reasonable spectral width.
Main Methods:
- Inverting the effective magnetic field midway through the adiabatic pulse sequence.
- Designing and simulating the modified RF pulse profiles.
Main Results:
- The converted pulses effectively act as 180-degree refocusing pulses.
- The modified pulses demonstrate robustness against significant changes in RF field strength.
- The pulses operate effectively over a practical spectral width.
Conclusions:
- Adiabatic inversion pulses can be readily adapted into robust refocusing pulses.
- These novel pulses offer improved performance for MRI techniques, particularly those using surface coils.