Rapid, Safe, and Simple Manual Bedside Nucleic Acid Extraction for the Detection of Virus in Whole Blood Samples
Published on: June 30, 2018
A comparison of SARS-CoV-2 RNA extraction with the QuickGene-810 Nucleic Acid Isolation System compared to the EZ1
Thuy Phan1, Rebecca Stephenson1, Tina Cai1
1NSW Health Pathology, Department of Microbiology and Infectious Diseases, Concord Repatriation and General Hospital, Hospital Rd, Concord, NSW 2139, Australia.
Abstract:
The QuickGene-810 Nucleic Acid Isolation System is a semi-automated extraction platform which may be used for RNA extraction. New methods were required to support the rapid increase in respiratory virus testing during the SARS-CoV-2 pandemic. The aim of this study was to assess SARS-CoV-2 RNA extraction using the QuickGene-810 kit compared to the EZ1 Advanced Extraction Platform for use on the AusDiagnostics SARS-CoV-2, Influenza and RSV 8-well RT-PCR assay. Qualitative results from all clinical samples were concordant between the QuickGene-810 and the EZ1 extraction methods, demonstrating that the QuickGene-810 kit is suitable for use in pathogen diagnostics. However, there was an average difference of approximately two cycles between the cycle threshold (Ct) values for both SARS-CoV-2 targets, suggesting that the EZ1 kit yields a higher concentration of nucleic acid extract, possibly related to its use of carrier RNA and/or smaller elution volume, which infers the possibility of false negative results for samples with very low viral loads.