Development of a standardized adsorbable organofluorine screening method for wastewaters with detection by combustion
Jenifer L Jones1, S Rebekah Burket2, Adrian Hanley2
1Office of Research and Development, U.S. Environmental Protection Agency, Cincinnati, OH, USA. jones.jenifer@epa.gov.
Abstract:
Per- and polyfluoroalkyl substances (PFAS) are man-made organofluorine chemicals that can contaminate environmental waters and have gained worldwide attention over the past two decades. PFAS are most frequently detected by mass spectrometric targeted analysis methods which may not detect all the PFAS in samples. This report describes the investigation of adsorbable organofluorine (AOF) with detection by combustion ion chromatography (CIC) for detection of PFAS in surface waters and wastewaters that adsorb to granular activated carbon (GAC) with the recognition that this technique measures more than just PFAS. Overall mean recoveries of 77-120% were obtained in 17 of the 18 tested surface water and wastewater matrices spiked with perfluoropentane sulfonate (PFPeS) and 55-119% mean recoveries were obtained in 11 of the 12 surface water and wastewater matrices spiked with a PFAS mixture. Poor method performance (34-39% mean recoveries) was observed in landfill leachate wastewater. Method detection limits of 1.4-2.2 microg L-1 were achieved using 100 mL sample volumes adsorbed onto commercially available GAC. This report demonstrates that this AOF technique can be a useful screening tool for estimating organofluorine concentrations when PFAS contamination is suspected.
More Related Videos
12:05Preparation of Hydrophobic Metal-Organic Frameworks via Plasma Enhanced Chemical Vapor Deposition of Perfluoroalkanes for the Removal of Ammonia
Published on: October 10, 2013
09:46Reducing Willow Wood Fuel Emission by Low Temperature Microwave Assisted Hydrothermal Carbonization
Published on: May 19, 2019
Related Concept Videos
Gas Chromatography: Types of Detectors-II
Sample Preparation for Analysis: Advanced Techniques
Acid digestion with strong acids is commonly used to dissolve inorganic materials that are insoluble (do not dissolve) in water. This method can be useful for...
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Flame Photometry: Lab
Atomic Fluorescence Spectroscopy
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing...