HyperArcTM Dosimetric Validation for Multiple Targets Using Ionization Chamber and RT-100 Polymer Gel.
Lucia Zirone1, Elisa Bonanno1, Giuseppina Rita Borzì1
1Medical Physics Department, Humanitas Istituto Clinico Catanese, 95045 Catania, Italy.
Gels (Basel, Switzerland)
|August 25, 2022
Summary
Single-isocenter stereotactic radiosurgery (SRS) for multiple brain metastases reduces treatment time. End-to-end tests using phantoms and polymer gel confirm high accuracy, validating its clinical use.
Area of Science:
- Radiotherapy Physics
- Medical Imaging
- Radiation Oncology
Background:
- Single-isocenter stereotactic radiosurgery (SRS) is emerging for treating multiple brain metastases.
- Pre-clinical end-to-end testing is crucial before implementing new radiotherapy techniques into clinical practice.
Purpose of the Study:
- To validate the accuracy of HyperArcTM single-isocenter SRS treatment plans for multiple brain metastases.
- To assess the feasibility of using polymer gel dosimetry and magnetic resonance imaging (MRI) for end-to-end verification.
Main Methods:
- Five HyperArcTM treatment plans were evaluated using ionization chambers and polymer gel dosimetry.
- End-to-end tests were conducted in a Mobius Verification Phantom (MVP) and a 3D-printed PseudoPatient® (PP) head phantom.
- Dose distributions were measured using 1.5T MRI and analyzed with gamma index criteria (e.g., 5% 2 mm, 3% 2 mm).
Main Results:
- Mean point dose differences were minimal: 1.01% in MVP and 0.23% in PP phantom.
- Gamma index passing rates exceeded 95% for both 2D and 3D dose distributions across tested criteria.
- High agreement was observed between calculated and measured dose distributions.
Conclusions:
- Single-isocenter SRS effectively reduces treatment time for multiple brain metastases.
- The study confirms the accuracy and reliability of this technique, even for targets distant from the isocenter.
- Polymer gel dosimetry with MRI provides a robust method for end-to-end verification of complex SRS plans.