Ex-Vivo Evaluation of "First Tip Closing" Radiofrequency Vessel Sealing Devices for Swine Small Intestinal
Luca Lacitignola1, Alberto Crovace1, Giuseppe Passantino2
1Department of Emergencies and Organ Transplantation, Section of Veterinary Clinics and Animal Production, University of Bari, Valenzano, 70010 Bari, Italy.
Abstract:
This study compared burst pressure (BP), number of activations, and histological assessment of ex vivo swine small intestine loops transected by stapler, a single fulcrum radiofrequency vessel sealing (RFVS) device, and the newly-developed jaws RFVS. Fifty (n = 50) 20 cm long jejunal loops were randomly assigned to be transected with RFVS devices and linear stapler (Caiman5, Caiman Maryland, Caiman12, Ligasure Atlas, and Stapler group as control respectively). Caiman5, Caiman12 and stapler required only one activation to complete the sealing. The mean BP in Caiman5 and Caiman Maryland groups were significantly lower (p < 0.05) than the S group as control and the other RFVS devices studied. RFVS Caiman12 and Ligasure Atlas produced mean BP values that were close to the Control and did not differ between them. The lumen was totally closed in the Caiman12 and Ligasure Atlas groups. The findings of this investigation were promising; we discovered that Caiman12 and Ligasure Atlas produce comparable mechanical capabilities as well as stapled intestinal closure, however Caiman12 need a single activation to complete the transection.
More Related Videos
11:17Surgical Technique for the Implantation of Tissue Engineered Vascular Grafts and Subsequent In Vivo Monitoring
Published on: April 3, 2015
06:39Novel Percutaneous Approach for Deployment of 3D Printed Coronary Stenosis Implants in Swine Models of Ischemic Heart Disease
Published on: February 18, 2020