Related Experiment Video
Updated: Aug 31, 2025

Magnetic Resonance Imaging of Multiple Sclerosis at 7.0 Tesla
Published on: February 19, 2021
Magnetic resonance imaging in patients with temporary external pacemakers
Daniel Benjamin Fyenbo1,2, Morten Steen Kvistholm Jensen1,2, Mads Brix Kronborg1,2
1Department of Cardiology, Aarhus University Hospital, 8200 Aarhus N, Denmark.
Aims:
To describe safety and feasibility of magnetic resonance imaging (MRI) in patients with transvenous temporary external pacemakers and whether artefacts affect the diagnostic image quality during cardiac MRI.
Methods And Results:
We reviewed records of all patients treated with temporary external pacing between 2016 and 2020 at a tertiary centre. Temporary pacing was established using a transvenous standard active fixation pacing lead inserted percutaneously and connected to a MRI-conditional pacemaker taped to the skin. All patients undergoing cardiac or non-cardiac MRI during temporary transvenous pacing were identified. Before MRI, devices were programmed according to guidelines for permanent pacemakers, and patients were monitored with continuous electrocardiogram during MRI. Of 827 consecutive patients receiving a temporary external pacemaker, a total of 44 (5%) patients underwent MRI (mean age 71 years, 13 [30%] females). Cardiac MRI was performed in 22 (50%) patients, while MRI of cerebrum, spine, and other regions was performed in the remaining patients. Median time from implantation of the temporary device to MRI was 6 (3-11) days. During MRI, we observed no device-related malfunction or arrhythmia. Nor did we detect any change in lead sensing, impedance, or pacing threshold. We observed no artefacts from the lead or pacemaker compromising the diagnostic image quality of cardiac MRI. MRI provided information to guide the clinical management in all cases.
Conclusion:
MRI is feasible and safe in patients with temporary external pacing established with a regular MRI-conditional pacemaker and a standard active fixation lead. No artefacts compromised the diagnostic image quality.
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies IV: Magnetic Resonance Imaging