2-Dimensional IMRT dose audit: An Indonesian multicenter study.
Ida Ayu Putu Inten Gayatri1, Andrian Dede Handika2, Wahyu Edy Wibowo3
1Department of Physics, Faculty of Mathematics and Natural Sciences, Universitas Indonesia, Depok, Indonesia; Department of Radiation Oncology, MRCCC Siloam Hospitals, Jakarta, Indonesia.
This study audited Indonesian radiotherapy centers using Intensity Modulated Radiation Therapy (IMRT). Most centers met dosimetry standards, showing IMRT accuracy, though stricter criteria revealed minor deviations.
Area of Science:
- Medical Physics
- Radiation Oncology
- Radiotherapy Quality Assurance
Background:
- Intensity Modulated Radiation Therapy (IMRT) is a precise radiation delivery technique.
- Accurate multileaf collimator function is critical for IMRT dose distribution.
- Deviations between planned and delivered IMRT doses can occur.
Purpose of the Study:
- To conduct a multicenter planar dosimetry audit of Indonesian radiotherapy centers.
- To evaluate IMRT delivery accuracy using AAPM TG-119 phantom data.
- To assess compliance with established dosimetry action levels.
Main Methods:
- Utilized AAPM TG-119 structure sets for dosimetry audits.
- Employed gamma index analysis with 3%/3 mm and 3%/2 mm criteria.
- Performed multicenter planar dosimetry evaluations across Indonesian radiotherapy centers.
Main Results:
- Achieved 100% gamma index passing rates for the 3%/3 mm criteria in most evaluations.
- Observed a decrease in gamma index passing rates with the stricter 3%/2 mm criteria.
- Confirmed that most participating centers met the tolerance limits for dosimetry action levels.
Conclusions:
- The multicenter audit demonstrated generally acceptable IMRT dosimetry accuracy in Indonesia.
- Stricter evaluation criteria highlight areas for potential improvement in IMRT delivery.
- This study provides a foundation for future standardized multicenter IMRT audits.