Is Highly Cross-Linked Polyethylene Safe for Use in High-Flexion Posterior Stabilized Total Knee Arthroplasty?
Young-Hoo Kim1, Jang-Won Park2, Young-Soo Jang1
1The Joint Replacement Center of Seoul Metropolitan SeoNam Hospital, Seoul, Republic of Korea.
Background:
Application of highly cross-linked polyethylene (HXLPE) to a posterior cruciate-substituting total knee arthroplasty (TKA) might add the risk of fracture and failure of the tibial polyethylene insert. The purpose of this study is to evaluate the long-term (up to 19 years) clinical and radiographic results of posterior cruciate-substituting TKAs with HXLPE or conventional polyethylene.
Methods:
This study analyzed the results of 1,217 patients (444 men and 773 women; mean age of 65 ± 7 years, range, 31-85) (2,434 knees) who had received a NexGen LPS-Flex prosthesis with a conventional tibial insert in one knee and the same prosthesis with an HXLPE tibial insert in the contralateral knee. The mean duration of follow-up was 17 years (range, 15-19).
Results:
The 2 groups did not differ significantly (P > .05) with regard to the clinical and radiographic results. No knee in either group had a fracture of the tibial polyethylene post or failure of the locking mechanism of the tibial polyethylene insert or osteolysis. Twenty-eight knees (2.3%) in the HXLPE group and 26 knees (2.1%) in the conventional polyethylene group were revised. The estimated survival rate at 17 years was 97.7% in the HXLPE group and 97.9% in the conventional polyethylene group.
Conclusion:
The data suggest that clinical and radiographic findings at a mean of 17 years after posterior cruciate-substituting TKA are the same for patients treated with HXLPE and those treated with conventional polyethylene.