Related Experiment Video
Updated: Aug 30, 2025

Multi-Scale Modification of Metallic Implants With Pore Gradients, Polyelectrolytes and Their Indirect Monitoring In vivo
Published on: July 1, 2013
Stress-dependent design and optimization methodology of gradient porous implant and application in femoral stem
Changning Sun1,2, Jianfeng Kang3, Ling Wang1,2
1State Key Laboratory for Manufacturing System Engineering, School of Mechanical Engineering, Xi'an Jiaotong University, Xi'an, ShaanXi, China.
Abstract:
Gradient porous structure made by additive manufacturing (AM) technology is potential to improve the long-term stability of orthopaedic implants through bone ingrowth while maintaining mechanical safety. In this study, a parametrical optimization methodology for the customized gradient porous implants was developed based on a stress-dependent design algorithm. Clinical requirements and manufacturing capabilities of AM were considered in the design procedure. A femoral stem with a minimum bone loss proportion of 2.4% by optimizing the control parameters. This study provided a feasible and flexible design approach for the customized implant with gradient porous structure or material components.