Calculation of scan length and size-specific dose at longitudinal positions of body CT scans using dose equilibrium
Xinhua Li1, Theodore A Marschall1, Kai Yang1
1Department of Radiology, Massachusetts General Hospital, Harvard Medical School, Boston, Massachusetts, USA.
Background:
Dose evaluation at longitudinal positions of body computed tomography (CT) scans is useful for CT quality assurance programs and patient organ dose evaluation. Accurate estimates depend on both patient size and scan length.
Purpose:
To propose practical evaluation of the average dose to the transverse slab of an axial image slice for adult body CT examinations, considering not only patient size but also scan length, and to compare the results with those of Monte Carlo (Geant4) simulation [Dsim (z)] and size-specific dose estimates at longitudinal positions of scans [SSDE(z)] from international standards (IEC publication no. 62985).
Methods:
In a scan series, the total dose at each z-axis location was calculated using the input information identical to the SSDE(z) evaluation. Each axial image slice (slice thickness, 2.5 or 5 mm) was first considered independently. Its z-axis coverage and CTDIvol (from the DICOM headers) were used to directly calculate a z-axis dose profile for the average dose over the cross-section of a water phantom, using the approach to equilibrium function. The phantom diameter was taken to be equal to the patient water equivalent diameter at that slice. The above was repeated at all slices and the dose at each z-axis location was accumulated from all profiles, referred to as Dcalc (z). For validation, we considered a cohort of 65 patients, who underwent chest and abdominopelvic examinations. The resultant Dcalc (z) was compared with Dsim (z) and SSDE(z), both available in a previous paper.
Results:
Dcalc (z) evaluation could be used to accurately assess the scan range average dose, with an accuracy of 7.1%-8.7% for 65 patients in two examinations. On individual image slices, the maximum difference in magnitude between Dcalc (z) and Dsim (z) [and between SSDE(z) and Dsim (z) in parentheses] was 37.5% (85%) [two edges (2 × 5 cm) of chest scan range], 17.8% (35.2%) (the remaining central region of chest scan), 26.8% (74.1%) [two edges (2 × 5 cm) of abdominopelvic scan range], and 14.2% (22.5%) (the remaining central region of abdominopelvic scan).
Conclusions:
Identical input data are used for Dcalc (z) and SSDE(z) evaluations. The latter is limited to the z-axis locations within scan range. At each image slice, SSDE(z) is equivalent to the midpoint dose of a fixed-mA scan of 15-30 cm (scan length). In contrast, Dcalc (z) considers dose accumulation from varying scan length (from sub-centimeter to about 1 m) and tube current, and dose profile is also computed outside scan range. Besides greatly improving dose evaluation for individual image slices, Dcalc (z) allows for evaluating dose accumulation from multiple series, which typically span different scan ranges. Our proposal may assist CT manufacturers and dose index monitoring software in assessing dose at longitudinal positions of body CT scans.
More Related Videos
07:31Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
Related Concept Videos
Biological Effects of Radiation
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET