Evaluation of the dosimetry and centralization of scout-view function in CBCT
Danúsia da Silva Vilela1, Luiz Roberto Coutinho Manhães Júnior1, Monikelly do Carmo Chagas Nascimento1
1Division of Oral Radiology, Faculdade São Leopoldo Mandic, Campinas, São Paulo, Brazil.
Abstract:
This study evaluated the centralization of the region of interest (ROI) in acquisition of the CBCT images, when the freely positionable scout-view (SV) function is applied. Additionally, the dosimetry of the acquired images was assessed in the SV function alone as well as in complete tomographic image in two different fields of view (FOV) (50x50 and 78x150mm). A three-location device was created to accommodate the dosimeters and the specimens, in the right, middle and left location during image acquisition. For dose assessment, thermoluminescent dosimeters were irradiated within the FOV and analyzed in a portable reader. For ROI evaluation, three specimens of gutta-percha stick were placed on the same device and the CT scans were acquired (CBCT OP 300 Maxio device, 90kV, 13mA, 85 µm voxel size, FOV of 50X50mm), with and without the SV, in three positions (3-9, 1-7 and 5-11 o'clock), simulating different regions of the mouth. Two image evaluations were performed, an objective and subjective. There was a slight percentage increase (1.36% to 1.40%) of the radiation dose with the use of SV. The distances were significantly greater in the images acquired without SV (p < 0.05). Every image obtained with SV was classified as being at the FOV's center. In conclusion, the results demonstrated that SVs function is effective to centralize the ROI in the FOV, increasing the scan precision and avoiding repetitions due to positioning errors.
More Related Videos
05:49Author Spotlight: Advancing CBCT and Digital Dental Image Integration with AI-Assisted Digitization
Published on: February 23, 2024
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography