Total Analytical Error and Measurement Uncertainty for Analytical Performance Evaluation and Determination of Gray
1Consultant on Biochemistry, Eskişehir, Turkey.
Objective:
Total analytical error (TAE) and measurement uncertainty (MU) are important approaches to evaluating and improving the quality of measurement procedures. This study evaluates glucose analytical performance (AP) according to TAE and MU and calculates gray zones of glucose critical value limits.
Methods:
Using TAE and MU values, AP was evaluated according to 5 different analytical performance specifications (APS) and the gray zones of critical value limits were calculated. The number of patients in these zones was compared.
Results:
TAE was higher than MU at all 3 levels. The AP for the low glucose level was poor. The number of patients in the gray zones was statistically higher in the TAE groups than in the MU groups (P < .05).
Conclusion:
TAE and MU values can be used to evaluate the AP of glucose measurement as well as to evaluate the compliance of patient results with decision limits by creating gray zones.
Related Concept Videos
Uncertainty in Measurement: Accuracy and Precision
Errors occurring during blood pressure monitoring
Several factors...
Uncertainty in Measurement: Reading Instruments
Accuracy and Errors in Hypothesis Testing
In hypothesis testing, the probability of making a Type I error, denoted as α, is commonly set at 0.05. This significance level indicates a 5%...
Systematic Error: Methodological and Sampling Errors
Sampling errors originate from improper sampling methods or the wrong sample population. These errors can be minimized by refining the sampling strategy. Defective instruments or faulty calibrations are the sources of instrumental...
Random and Systematic Errors