Novel Outcome Analysis Tool for Hypoglossal Nerve Stimulator Sensor Lead Function and Comparison by Incision Type (2
Mohamad Z Saltagi1, Kayla Powell1, Abdul K Saltagi1
1Department of Otolaryngology-Head and Neck Surgery, Indiana University School of Medicine, Indianapolis, Indiana, U.S.A.
The Laryngoscope
|September 2, 2022
Summary
This study introduces new ways to measure hypoglossal nerve stimulator (HNS) sensor lead function. The 2-incision technique for HNS placement showed better waveform syncing and overall impression compared to the 3-incision method.
Area of Science:
- Medical Devices
- Neurosurgery
- Sleep Medicine
Background:
- No established outcome measures exist for evaluating hypoglossal nerve stimulator (HNS) sensor lead function.
- Developing reliable methods to assess HNS performance is crucial for patient care and device optimization.
Purpose of the Study:
- To develop novel functional outcome measures for intraoperative sensor electrode function in HNS.
- To compare the outcomes of 2-incision versus 3-incision techniques for HNS placement based on these new measures.
Main Methods:
- Retrospective cohort study of 100 HNS patients.
- Development of structured parameters to analyze intraoperative waveforms (syncing, amplitude, leakage, morphology, artifact, overall impression).
- Blinded assessment of waveforms by sleep surgeons and specialists using a Likert scale.
Main Results:
- No significant differences in patient demographics or major complications between 2-incision and 3-incision groups.
- High interrater reliability (>0.7 ICC) for the developed outcome measures.
- The 2-incision technique demonstrated statistically better waveform syncing and overall impression compared to the 3-incision technique.
Conclusions:
- This study presents the first structured method for measuring HNS sensor electrode outcomes with consistent interrater reliability.
- The findings support the adoption of the 2-incision technique for HNS sensor lead placement.
- The developed outcome measures represent a valuable tool for clinical and research settings, requiring further validation.