Effects of Beamforming Techniques on Quality of Ultrasound Computed Tomography Images
Razieh Solgi1,2, Hossein Ghadiri3,4
1MSc, Department of Medical Physics and Biomedical Engineering, Tehran University of Medical Sciences, Tehran, Iran.
Background:
In body tissues, tumors generally have different speeds of sound (SOS) than normal tissues. In this respect, ultrasound computed tomography (UCT) can generate a cross-sectional SOS map as an innovative ultrasound imaging method. This technique can produce images with a resolution of millimeters and a high signal-to-noise ratio.
Objective:
This study aimed to improve UCT image quality without increasing breast cancer screening and diagnosis time.
Material And Methods:
In this analytical study, a ring-shaped UCT breast imaging system was simulated using the K-wave toolbox of MATLAB. The system has a 20 cm diameter and 256 ultrasonic piezoelectrics placed in the ring's circumference. Different beamforming techniques imaged two designed phantoms (i.e., resolution and contrast), and the resolution and contrast to noise ratio (CNR) were calculated.
Results:
The results of resolution phantom imaging without any beamforming showed that only bars with the value of 0.125 and 0.167 lp/mm were distinguishable, and the 0.1 bars were not recognizable in the imaging. In addition, increasing the number of transmitters led to no noticeable change in resolution for 0.125 and 0.167 lp/mm bars. In all beamforming techniques for imaging the contrast phantom, the CNR parameter up to an object with a diameter of 8 mm increases with increasing diameter without any change.
Conclusion:
The beamforming technique using three simultaneous transmitters improved the resolution by about 1 mm compared to the normal strategy. In addition to high-contrast images, beamforming with 9 simultaneous transmitters led to a preferable technique.
More Related Videos
07:38Real-time Monitoring of High Intensity Focused Ultrasound HIFU Ablation of In Vitro Canine Livers Using Harmonic Motion Imaging for Focused Ultrasound HMIFU
Published on: November 3, 2015
16:01An Experimental Protocol for Assessing the Performance of New Ultrasound Probes Based on CMUT Technology in Application to Brain Imaging
Published on: September 24, 2017
Related Concept Videos
Ultrasonography
During an ultrasonography procedure, a handheld device called...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies II: Ultrasonography