Calvarial Model of Bone Augmentation in Rabbit for Assessment of Bone Growth and Neovascularization in Bone Substitution Materials
Published on: August 13, 2019
8.4K
Bionate Biocompatibility: In Vivo Study in Rabbits
Amparo Vanaclocha-Saiz1, Vicente Vanaclocha2, Carlos Atienza1
1Instituto de Biomecánica (IBV), Universitat Politècnica de Valencia, Valencia 46022, Spain.
ACS Omega
|September 5, 2022
Summary
Bionate 80A spinal implants in rabbits showed no significant toxicity or migration. Microscopic analysis revealed a localized tissue response outside the dura mater, with no adverse effects on organs or lymph nodes.
Area of Science:
- Biomaterials Science
- Toxicology
- Spinal Surgery
Background:
- Foreign material implantation can cause adverse tissue reactions, including inflammation, osteolysis, and migration.
- Previous human explants of Bionate 80A showed minor wear but limited data on systemic or spinal cord responses.
- Understanding the biological response to Bionate 80A in the spinal epidural space is crucial for its clinical application.
Purpose of the Study:
- To evaluate the local and systemic biological response to Bionate 80A implanted in the rabbit spinal epidural space.
- To assess potential toxicity, inflammation, and migration of Bionate 80A particles.
- To determine the safety profile of Bionate 80A in a spinal implantation model.
Main Methods:
- Bionate 80A was implanted into the spinal epidural space of 12 New Zealand white rabbits; 12 served as controls.
- Animals were monitored for clinical parameters and weight changes over 6 months.
- Histopathological examination of tissues, blood biochemistry, urine tests, and necropsy were performed to assess reactions and damage markers.
Main Results:
- No statistically significant differences in clinical parameters or weight were observed between groups.
- Minor, non-significant changes in basophils and eosinophils were noted in the implant group.
- Microscopic examination revealed a localized chronic granulomatous response with giant multinucleated cells containing Bionate particles outside the dura mater; no particle migration to lymph nodes or organs was detected.
Conclusions:
- Bionate 80A implantation in the rabbit lumbar epidural space did not induce significant systemic toxicity or inflammatory changes.
- A localized foreign body reaction occurred, confined to the tissues surrounding the implant and outside the dura mater.
- Bionate 80A particles demonstrated no migration across the dura mater or to distant lymph nodes and organs in this rabbit model.