CONTRIBUTION OF ORGAN-BASED TUBE CURRENT MODULATION TO THE REDUCTION OF LENS EXPOSURE DOSE IN HEAD 4D CT IMAGING: A
Keisuke Nagamoto1, Ryo Watanabe1, Tomonori Kawachino1
1Department of Radiology, Hospital of the University of Occupational and Environmental Health, Japan, 1-1 Iseigaoka, Yahatanishi-Ku, Kitakyushu, Fukuoka, Japan.
Abstract:
The purpose of this study was to evaluate the effectiveness of organ effect modulation (OEM) in reducing the lens dose in 4D computed tomography (CT) of the head in volume-acquisition (NVA) mode. Six radiophotoluminescent dosemeters were placed on the head of a RANDO phantom. The doses absorbed by the organs and image noise change rate were determined. The lens doses without OEM (i.e. in the OEMoff case) were higher than those with the same target standard deviation and volume-computed tomography dose index (CTDIvol) as in the OEMoff case (p < 0.01). The image noise change rate was 11%. OEM reduced the lens dose during head 4D CT imaging in the NVA mode by 18%. Furthermore, the feasibility of lens dose reduction while ensuring sufficient image quality was confirmed under the condition in which OEM was employed with the same CTDIvol as in the OEMoff case.
More Related Videos
10:23Author Spotlight: A Machine-Vision Approach to Transmission Electron Microscopy Workflows, Results Analysis and Data Management
Published on: June 23, 2023
08:54Treatment of Liver Metastases Using an Internal Target Volume Method for Stereotactic Body Radiotherapy
Published on: May 8, 2018
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System V: CT