Improving Performance and Access to Difficult-to-Reach Anatomy with a Powered Articulating Stapler.
Zhifan F Huang1, James A Vandewalle1, Jeffrey W Clymer1
1Ethicon, Inc, Cincinnati, OH, USA.
Medical Devices (Auckland, N.Z.)
|September 9, 2022
Summary
The Echelon 3000 Stapler offers better surgical precision and tissue handling. This new surgical stapler demonstrated superior performance in critical areas like seal strength and staple formation in thick tissues.
Area of Science:
- Surgical Technology
- Biomedical Engineering
- Surgical Oncology
Background:
- Modern surgical staplers require precise placement and transection, especially in challenging anatomical spaces and with thick tissues.
- The Echelon 3000 Stapler (ECH3) was developed to accommodate tissue and anatomical variations, improving surgical workflow.
- This study evaluated the ECH3 against a comparator stapler for access, seal strength, staple formation, and end effector stability.
Purpose of the Study:
- To evaluate the performance characteristics of the new Echelon 3000 Stapler (ECH3).
- To compare the ECH3's ability to complete a Low Anterior Resection (LAR) transection in one or two firings against a marketed stapler.
- To assess seal strength, staple formation in thick tissue, and end effector stability of the ECH3.
Main Methods:
- Virtual modeling of pelvic anatomy for Low Anterior Resection (LAR) and Monte Carlo simulations to assess stapler firing probability.
- Ex vivo testing using stapled porcine ileum to measure leak pressure and leak rates under critical pressures.
- Assessment of malformed staple rates in thick tissues (3.3- and 4.0-mm) and comparison of end effector stability during firing and under load.
- Surgeon surveys on device functionality and ease of use post-preclinical evaluation.
Main Results:
- The ECH3 demonstrated a significantly higher probability of completing an LAR transection in one or two firings compared to the comparator stapler.
- Stapled porcine ileum with the ECH3 showed a higher median initial leak pressure and a lower rate of leaks at critical pressures (40 and 50 mmHg).
- The ECH3 produced fewer malformed staples in thick tissues and exhibited less end effector angular movement and deflection during firing and under load.
- Surgeons reported precise placement and easy one-handed operation with the ECH3.
Conclusions:
- The Echelon 3000 Stapler offers improved access, tighter seals, reduced malformed staples, and enhanced end effector stability.
- These performance advantages were recognized by surgeons during preclinical evaluations.
- The ECH3 represents a significant advancement in surgical stapler technology for complex procedures.